Remanufactured Sakura Tissue-Tek Prisma Staining System with 4740 Coverslipper
Bring automated precision to your histology workflow with the remanufactured Sakura Tissue-Tek Prisma Staining System, paired with the Tissue-Tek Film 4740 Automated Coverslipper for end-to-end slide preparation.
Designed for high-throughput surgical pathology and clinical laboratory environments, this remanufactured staining system takes slides seamlessly from the staining process through coverslipping in a single, continuous workflow. Whether you're running routine H&E stain protocols or more complex staining procedures, this Sakura Tissue-Tek Prisma staining system delivers the consistent, reliable results your lab demands.
Key Features:
-
Integrated automated slide stainer and Film coverslipper for end-to-end slide preparation
-
The Tissue-Tek Film 4740 Coverslipper links directly to the Prisma stainer via optional transfer link station, enabling automatic slide delivery from staining to coverslipping
-
Coverslipping performed at approximately 3 seconds per slide with continuous processing of up to 3 pre-loaded baskets (60 slides)
-
Maximum throughput of 1,080 coverslipped slides per hour
-
Resin-coated Tissue-Tek Film and xylene system eliminates the need for cover glass and mounting media
-
User-selectable film lengths: 45 mm, 50 mm, 55 mm, and 60 mm
-
Microcomputer-controlled mechanical movements with four-line LCD display for real-time status, setup menus, error codes, and messages
-
Built-in carbon filtration system absorbs xylene fumes, with optional duct connection adapter for external venting
-
12-position carousel unloading unit holds up to 240 coverslipped slides
-
Compatible with Tissue-Tek DRS 2000 and Prisma slide baskets
Applications:
-
Routine H&E staining and automated coverslipping for surgical pathology
-
High-throughput histology and clinical laboratory slide preparation
-
Cytology and hematology specimen processing
-
Continuous staining workflows requiring seamless integration between stainer and coverslipper
-
Remanufactured staining system for labs seeking cost-effective access to the Sakura Tissue-Tek Prisma platform
Specifications:
|
Parameter |
Specification |
|
Processing Speed |
Up to 1,080 slides per hour; ~3 seconds per slide |
|
Loading Capacity |
3 baskets (60 slides) simultaneously |
|
Unloading Capacity |
12-basket carousel unit (240 coverslipped slides) |
|
Acceptable Slide Dimensions |
24.7–26.5 mm (W) x 74.7–76.5 mm (L); 0.9–1.2 mm thickness |
|
Film Width |
24 mm |
|
Film Lengths (selectable) |
45 mm, 50 mm, 55 mm, 60 mm |
|
Film Material |
Cellulose Triacetate with Resin and Back Coatings |
|
Compatible Solvent |
Reagent grade or analytical grade xylene only |
|
Xylene Bottle Capacity |
500 ml |
|
Waste Bottle Capacity |
200 ml |
|
Xylene Dispense Volume |
5 user-selectable levels (1 min – 5 max) |
|
Fume Control |
Dual activated carbon cartridge filtration; optional external duct port (38 mm or 75 mm OD) |
|
Display |
Four-line LCD with nine keypad switches and three LEDs |
|
Coverslipper Dimensions |
72 cm (W) x 58 cm (D) x 69 cm (H) / 28 in (W) x 23 in (D) x 27 in (H) |
|
Instrument Weight |
75 kg (165 lbs) |
|
Operating Temperature |
10°C to 40°C (50°F to 104°F) |
|
Operating Humidity |
30–85% relative humidity (non-condensing) |
|
Atmospheric Pressure |
70 kPa to 106 kPa |
|
Storage Temperature |
-10°C to 60°C |
|
Storage Humidity |
30–95% relative humidity (non-condensing) |
|
Safety Standards |
IEC 61010-1 2nd Ed., CAN/CSA C22.2 No. 61010-1 2nd Ed., UL 61010-1 2nd Ed.; CE registered |
|
Model |
4740 (USA) |
Electrical Specifications:
|
Parameter |
Specification |
|
Voltage |
115 VAC ±10% |
|
Frequency |
60 Hz |
|
Amperage |
1.3 A |
|
Max Current Draw |
<15 A |
Maintenance:
Proper maintenance ensures reliable performance and long instrument life. Key maintenance intervals include:
-
Daily: Empty the xylene waste bottle; inspect and clean internal components as needed
-
Every 2 weeks: Replace activated carbon filters after two weeks of continuous use
-
Monthly: Empty and clean the film waste tray; brush rollers and blade assembly to remove resin dust
-
Every 10 rolls of film (~10,000 slides): Replace the cutter blade, or sooner if film cuts become ragged or torn
-
As needed: Soak slide baskets in xylene to remove resin/xylene residue buildup; inspect baskets weekly for wear or stress
-
Extended idle periods: Power down the instrument and roll the film back onto the roll to prevent curvature from roller contact
Warranty
This remanufactured Sakura Tissue-Tek Prisma staining system is sold by Medical Equipment Source and is backed by a 1-year warranty from MES. Please contact Medical Equipment Source for full warranty terms and service details.
Optimize Your Workflow Today!
The remanufactured Sakura Tissue-Tek Prisma Staining System with 4740 Film Coverslipper is a complete, integrated solution for laboratories seeking automated, high-throughput slide staining and coverslipping in a single system. From routine H&E staining to continuous surgical pathology workflows, this staining system reduces manual handling, increases throughput, and delivers consistently coverslipped slides ready for diagnosis.
Contact Medical Equipment Source today to request a quote and find out how this staining system can elevate efficiency in your histology or pathology lab.
Documents & Manuals
Featured Products
Protect Your Lab with Preventative Maintenance
Preventative maintenance helps reduce unexpected downtime, support regulatory compliance, and keep lab equipment performing reliably. MES works with your team to build maintenance plans based on your instruments, usage, and operational needs.
Get Expert Guidance for Your Lab
Whether you need service, equipment, or planning support, our team is ready to help you find the right solution for your lab.